Flame Photometer Microprocessor Based
MODE |
General | Serum | Urine |
| Na: 0-100 ppm, Up to 250 meq/L, 1 : 100 dil |
Na: 0-200 ppm 1:100 dil |
Na:0-250 ppm 1:100 dil |
|
| K: 0-100 ppm Up to 250 meq/L, 1 : 100 dil |
K: 0-10 ppm 1:100 dil |
K: 0-100 ppm 1:100 dil |
|
| Ca: 0-100 ppm Up to 250 meq/L, 1 : 100 dil |
— |
Ca: 0-100 ppm 1:2 dil |
|
| Li: 0-100 ppm Up to 250 meq/L, 1 : 100 dil |
Li: 0-2 ppm 1:5 dil |
— |
|
| Sensitivity | General: Na : 0.5 ppm, K : 0.5 ppm, Li : 0.5 ppm, Ca: 15 ppm | ||
| Filters | Na & K, | Ca & Li (Optional if Ordered) | |
| Resolution | 0.1 ppm/ meq | ||
| Reproducibility | < 1% coefficient of variation for 20 consecutive samples using 10 ppm, Na set as maximum standard. | ||
| Curve Fit Accuracy | < 2% error when 3ppm Na/K and 5ppm Li, set as max.Standard. | ||
| Display | 128 X 64 Graphical LCD | ||
| Average Time | In built Software | ||
| Flame System | LPG and Dry Oil free Air | ||
| Detector | Highly Sensitive Silicon Photo Diode | ||
| Calibration | Up to 5 point calibration with curve fitting Software. | ||
| Gas Control | Adjustable with knob | ||
| Flame Failure | Auto Detection | ||
| Gas Cut Off | Gas feeding, auto cut off in case of Power Failure | ||
| Ignition System | Auto Ignition | ||
| RS-232 | Provision for Data Transfer to PC | ||
| Power | 230 V +- 10%, 50 Hz. AC | ||
| Dimension | 280X 250 X210 mm (L X W X H) | ||
| Weight | 7.4 Kg (approx.) | ||
| Compressor Unit | |||
| Air Supply | By oil free mini Compressor unit with pressure Regulator | ||
| Combustion | LPG controlled by precision Regulator | ||
| Power | 230 V +-10%, 50Hz AC | ||
| Dimension | 290 X 250 X 210 mm (L X W X H) | ||
| Weight | 8 Kg (approx.) | ||